    Title: Near infrared tomographic system based on high angular resolution mechanism - Design, calibration, and performance
    Authors: Pan,MC;Chen,CH;Pan,MC;Shyr,YM
    Contributors: 生物醫學工程研究所
    Keywords: DIFFUSE OPTICAL TOMOGRAPHY;IMAGE-RECONSTRUCTION;MODEL;INSTRUMENTATION;SCATTERING
    Date: 2009
    Issue Date: 2010-06-29 17:54:22 (UTC+8)
    Publisher: 中央大學
    Abstract: This study aims at the design, implementation, and verification of a near infrared (NIR) diffuse optical tomography (DOT) electro-optical imaging system based on a measuring instrument with single rotating source/detector scanning mechanism and an image reconstruction scheme implemented on an off-line PC. The developed measurement instrument possesses a high degree of angular resolution and is operated in a continuous-intensity scheme to acquire output NIR intensity signals. In this paper, some considerations of this measuring instrument are investigated such as phantom preparation, temporal stability, and container influence, etc. Experimental trials were conducted using both homogeneous and heterogeneous phantoms made of high-scattering microspheres or Intralipid. In order to verify the developed imaging system with tomographic image reconstruction, the intensity was acquired from the heterogeneous Intralipid phantom using NIR light at 800 nm, where the homogenous background as the normal tissue is inserted with an inclusion as the tumor. For the comparison of heterogeneous phantoms with homogeneous phantoms, 1-D measured intensity signals indicate the angular position of the inclusion, especially for the case of the inclusion near the phantom boundary. Furthermore, promising results of corresponding optical-property images can be reconstructed with the measured intensities as inputs of the image reconstruction scheme. Obviously, the developed NIR DOT tomographic system has been successfully built upon the experimental, analytical, and computational work. (C) 2008 Elsevier Ltd. All rights reserved.
    Relation: MEASUREMENT
    Appears in Collections:[Institute of Biomedical Engineering] journal & Dissertation

    Files in This Item:

    File Description SizeFormat
    index.html0KbHTML886View/Open


    All items in NCUIR are protected by copyright, with all rights reserved.

    社群 sharing

    ::: Copyright National Central University. | 國立中央大學圖書館版權所有 | 收藏本站 | 設為首頁 | 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::
    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 隱私權政策聲明