中大機構典藏-NCU Institutional Repository-提供博碩士論文、考古題、期刊論文、研究計畫等下載:Item 987654321/63455
English  |  正體中文  |  简体中文  |  Items with full text/Total items : 80990/80990 (100%)
Visitors : 41648444      Online Users : 1456
RC Version 7.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
Scope Tips:
  • please add "double quotation mark" for query phrases to get precise results
  • please goto advance search for comprehansive author search
  • Adv. Search
    HomeLoginUploadHelpAboutAdminister Goto mobile version


    Please use this identifier to cite or link to this item: http://ir.lib.ncu.edu.tw/handle/987654321/63455


    Title: C-arm影像導引系統應用於股骨幹骨折骨髓內釘遠端鎖定之臨床實作
    Authors: 吳鴻康;Wu,Hung-Kang
    Contributors: 生物醫學工程研究所
    Keywords: 手術導引;骨髓內釘;遠端固定;C-arm影像;surgical navigation;intramedullary nail;distal locking screw;C-arm image
    Date: 2014-01-27
    Issue Date: 2014-04-02 15:18:57 (UTC+8)
    Publisher: 國立中央大學
    Abstract: 股骨幹骨折是骨科常見的疾病,常因高能量的傷害如車禍,重物打擊所引起,主要的治療方式是以骨髓內釘來固定骨頭,但遠端鎖定式骨釘的置入常是手術者棘手的問題,傳統手術需要照射多次X光(C-arm),病人及醫療團隊都會被迫接受大量輻射, 本實驗希望經由電腦導引系統實施遠端鎖定式骨釘的手術置入,以期能更準確的打入內置物,並縮短時間及減少移動式X光機( C-arm )的照射。
    本實驗最後一共收集了12個病人,6位病人使用電腦導引手術,6位使用傳統手術,同一組醫療人員隨機使用電腦導引系統及傳統手術方式紀錄並比較其手術時間及C型臂之X光照射次數,實驗組手術時間47±14.17分鐘,C-arm照射次數為4.83±0.56次,對照組手術時間為45±21.67分鐘 , C-arm照射次數為40.67±27.22次。
    電腦導引與傳統手術比起來, 有明顯差異的還是在X光的照射次數上, 在手術時間上相對而言並無明顯差異, 在電腦導引所花的時間稍微多一點, 但可以發現時間似乎隨著使用的增加而減少。; Femoral shaft fractures are common orthopedic diseases , caused by high energy often as heavy combat damage or a car accident. The main treatment is to fix the bone by intramedullary nail , but the nail's distal locking screw insertion pass through the bone to distal locking hole is often thorny issues by surgeon. Traditional surgery requires several X-ray irradiation (C-arm), patients and the medical teamwork will be forced to accept large amounts of radiation. We hope to insert the distal locking screws smoothly via computer navigative surgery systems, in order to more accurately insertion by distal locking screws , shorten of the operative time and reduce irradiation by the times of the X ray.
    We collected a total of 12 patients , 6 patients using computer navigation surgery, 6 using traditional surgery , the same group of medical staff using a random study to compare computer navigation systems and traditional surgery. This study record and compare their operation time and the C-arm X-ray irradiation times , the experimental group's operation time was 47 ± 14.17 minutes, C-arm irradiation times of 4.83 ± 0.56 times, the control group's operative time was 45 ± 21.67 minutes, C-arm irradiation times of 40.67 ± 27.22 times.
    Computer navigation surgery compared with traditional C-arm surgery, there are still significant differences in the number of X-ray irradiation. There was no significant difference in operative time. The computer navigation spend a little more time , but seems to decrease with increasing use .
    Appears in Collections:[Institute of Biomedical Engineering] Electronic Thesis & Dissertation

    Files in This Item:

    File Description SizeFormat
    index.html0KbHTML897View/Open


    All items in NCUIR are protected by copyright, with all rights reserved.

    社群 sharing

    ::: Copyright National Central University. | 國立中央大學圖書館版權所有 | 收藏本站 | 設為首頁 | 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::
    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 隱私權政策聲明