中大機構典藏-NCU Institutional Repository-提供博碩士論文、考古題、期刊論文、研究計畫等下載:Item 987654321/54294
English  |  正體中文  |  简体中文  |  Items with full text/Total items : 80990/80990 (100%)
Visitors : 40251104      Online Users : 318
RC Version 7.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
Scope Tips:
  • please add "double quotation mark" for query phrases to get precise results
  • please goto advance search for comprehansive author search
  • Adv. Search
    HomeLoginUploadHelpAboutAdminister Goto mobile version


    Please use this identifier to cite or link to this item: http://ir.lib.ncu.edu.tw/handle/987654321/54294


    Title: 利用相位式表面電漿共振系統檢測免疫球蛋白鍵結之應用分析;Immunoglobulin Binding Analysis by Utilizing a Surface Plasmon Resonance Phase-Measurement System
    Authors: 林萱;Lin,Shiuan
    Contributors: 照明與顯示科技研究所
    Keywords: 表面電漿共振;生物分子;Surface Plasmon Resonance;Immunoglobulin
    Date: 2012-07-31
    Issue Date: 2012-09-11 18:43:19 (UTC+8)
    Publisher: 國立中央大學
    Abstract: 本研究之動態式表面電漿共振感測器是以相位移干涉為基礎,以偵測器擷取光強訊號,代入五步相移法進行解相。由實驗結果可得知系統於3600秒下相位穩定度為1.2214度、系統靈敏度為1.1465✕10^4 (Degree/RIU)、系統折射率解析度可達1.0653✕10^-4 (RIU)。利用鹽水實驗可得折射率與相位差關係,進而去推算未知鹽水溶液折射率,經折射率計量測與實驗後內插所得之折射率,兩者誤差約為0.023 %。於免疫球蛋白量測部分,利用濃度分別為200 μg/ml、100 μg/ml、50 μg/ml、25 μg/ml之Anti-IgG與固定於金膜上之IgG進行鍵結,所對應相位變化分別為:78.7476°、58.8533°、41.0506°、33.1399°,由相位與時間關係圖去計算結合速率常數k_a與解離速率常數k_d,其值分別為4.1139✕10^4M^-1s^-1與2.8768✕10^-4s^-1。經由鹽水實驗與生物分子鍵結量測可得知SPR相位訊號隨待測液濃度改變之關係式,故可利用此關係曲線來進行未知濃度檢測。A dynamic surface-plasmon-resonance (SPR) sensor was developed based on interferometric phase measurements. Temporal intensity signals were acquired by photodetectors and processed by the Schwider-Hariharan five-step algorithm to obtain the phases. The achieved phase-detection stability was 1.2214 degrees in a 3600-second period and the system sensitivity was 1.1465×10^4 degree/RIU (refractive index unit). The corresponding system resolution was 1.0653×10^-4 RIU. Salt-water mixture measurements were performed to characterize the relation between the measured phase difference and the refractive-index variation of the specimen. Several known concentrations of salt-water mixtures were flowed into the SPR system to measure the phases. A refractometer was utilized to measure the refractive indices. The results showed a linear correspondence between the phase difference and the refractive index. The measurement of an unknown salt-water mixture showed a relative refractive-index error of about 0.023 % between the refractometer measurement and the linear interpolation result from the SPR system.For the antibody-antigen binding experiments, the phase differences caused by the injection of anti-IgG with concentrations of 25 μg/ml, 50 μg/ml, 100 μg/ml and 200 μg/ml were measured to be approximately 33.1399°, 41.0506°, 58.8533°, and 78.7476°, respectively. From the relation between the time and phase difference, the association rate constant ka and the dissociation rate constant kd were calculated to be 4.1139×10^4 M^-1s^-1 and 2.8768×10^-4 s^-1, respectively. By utilizing the relation between the SPR signals and the concentrations of the analyte from salt-water mixture measurements and antibody-antigen binding experiments, the concentration of unknown solutions can be estimated by this SPR system.
    Appears in Collections:[Institute of Lighting and Display Science ] Electronic Thesis & Dissertation

    Files in This Item:

    File Description SizeFormat
    index.html0KbHTML902View/Open


    All items in NCUIR are protected by copyright, with all rights reserved.

    社群 sharing

    ::: Copyright National Central University. | 國立中央大學圖書館版權所有 | 收藏本站 | 設為首頁 | 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::