博碩士論文 965201014 完整後設資料紀錄

DC 欄位 語言
DC.contributor電機工程學系zh_TW
DC.creator洪君岳zh_TW
DC.creatorJyun-Yue Hongen_US
dc.date.accessioned2010-8-20T07:39:07Z
dc.date.available2010-8-20T07:39:07Z
dc.date.issued2010
dc.identifier.urihttp://ir.lib.ncu.edu.tw:88/thesis/view_etd.asp?URN=965201014
dc.contributor.department電機工程學系zh_TW
DC.description國立中央大學zh_TW
DC.descriptionNational Central Universityen_US
dc.description.abstract隨著醫學科學與超大型積體電路的快速發展,應用在植入人體內的醫用電子設備也不斷演進,發展至今已可被應用於調節生理功能、刺激神經、肌肉及治療多種疾病。 本論文的設計為一個應用於植入式視覺輔具的電刺激器。電刺激器必須透過電極作為介面才能對神經或肌肉進行電刺激的動作。然而電極-組織介面阻抗可能會因接觸不良、電極大小與材質的差異,又或者電極本體受刺激電流、環境等因素而產生變化,其變異範圍在10–100KΩ。當介面阻抗值過大時,刺激電流會受到供應電壓的限制,無法在刺激週期內達成有效的電刺激。 為了讓電刺激器能滿足阻抗變異的需求,根據阻抗變異與電刺激參數的設定,論文中提出了一個可產生三倍供應電壓輸出的高電壓輸出級驅動電路作為電壓控制電刺激器的想法,並以台積電0.18μm 互補式金氧半導體的標準製程實現。其好處是不需花費其他昂貴的特殊製程,又可以與其他電路整合於同一晶片。 在進行電刺激之前,使用阻抗量測電路測量介面阻抗,而數位脈波寬度調變電路可針對不同介面阻抗與刺激電荷量的需求,產生出適當的脈波寬度,使得高電壓輸出級驅動電路能有效的調整刺激電荷量,進而使得植入式電刺激器達到最佳的功效。 zh_TW
dc.description.abstractWith the rapid development of medical science and very large scale integration circuit, medical electronic devices which are used to implant in the human body have been evolved. Furthermore, some medical electronic devices have already been applied for regulating human’s physiological functions, stimulating nerves, muscles, and treatment of many diseases. This thesis aims to design an electrical stimulator for implanted visual prosthesis. The designed electrical stimulator stimulates nerves or muscles using electrodes as the interface. In general, the electrode-tissue interface impedance may changes in the rage of 10–100 KΩ due to poor contact, the electrode size and material differences, or electrode-self by stimulus current and environmental factors, and so on. Using impedance measurement circuit to measure interface impedance before stimulated, and this thesis proposes the digital pulse width modulator to produce proper pulse widths for different interface impedances, enabling electrical stimulator to adjust the stimulus charge, thus making electrical stimulator to achieve the best effect. However, the stimulus current is limited by the supply voltage, effective stimulating may be unable to be achieved in a stimulus period when the interface impedance is too large. In order for electrical stimulator to satisfy the requirement of impedance variation, this thesis also proposes a high voltage output driver, which can generate three times supply voltage output, for the voltage-controlled electrical stimulator. Moreover, such a high voltage output driver accommodating to impedance variation and stimulus parameter setting is implemented in TSMC 0.18μm standard CMOS technology to demonstrate the feasibility of the proposed electrical stimulator. An advantage of the designed high voltage output driver is that it can be fully integrated within other circuit without extra process costs. en_US
DC.subject數位脈波寬度調變電路zh_TW
DC.subject阻抗量測電路zh_TW
DC.subject電極組織介面zh_TW
DC.subject高電壓電路設計zh_TW
DC.subject植入式電刺激器zh_TW
DC.subjectdigital pulse width modulationen_US
DC.subjectimpedance measurement circuitsen_US
DC.subjectelectrode-tissue contacten_US
DC.subjecthigh voltage techniquesen_US
DC.subjectimplantable electrical simulatorsen_US
DC.title應用於植入式生醫系統的適應性高電壓電刺激器與阻抗量測電路之分析與設計zh_TW
dc.language.isozh-TWzh-TW
DC.titleAnalysis and Design of an Adaptive High-Voltage Electrical Stimulator and Impedance Measurement Circuitry for Implantable Biomedical Systemsen_US
DC.type博碩士論文zh_TW
DC.typethesisen_US
DC.publisherNational Central Universityen_US

若有論文相關問題,請聯絡國立中央大學圖書館推廣服務組 TEL:(03)422-7151轉57407,或E-mail聯絡  - 隱私權政策聲明