摘要(英) |
Since 1982, cancer has been the top ten cause of death among Chinese people. Today, radiation therapy is considered a viable non-invasive cancer treatment, so about 50% of cancer patients receive radiation therapy during treatment, and radiation therapy includes photon therapy and proton therapy. Recent advances in proton therapy have provided a highly uniform and high-dose-rate of precise treatment for the treatment of tumors. To achieve this goal, a better quality assurance program is needed to ensure treatment quality. Therefore, the starting point of the detector design discussed in this paper is the application of proton therapy.
The XY strip detector studied in this paper is characterized by a large area of induction receiving surface (345.44 × 345.44 mm2), high spatial resolution ( < 60μm), high dynamic response range (20 bits, ADC reachable 1,048,576) with high sensitivity (50 fC), and has flexible operation options such as integration time and charge accumulation range. In addition, the LabVIEW program′s human-machine interface is used to collect experimental data and monitor beam changes in real time. Therefore, the existing hardware structure, electronic components, and operation interface of the XY strip detector can meet the current hospital proton beam-related quality assurance operations.
The detector can be used to understand the performance of the detector when it is used in the daily inspection of the pen tip scanning beam parameters of the hospital. The overall resolution is better than that of the current hospital. In the proton beam patient dose test, the high signal dynamic range and signal change rate of the detector can clearly distinguish the absorbed dose distribution when the beam is incident with different energies if it can be supplemented with appropriate blocking power and simulation. Program to reconstruct the three-dimensional dose distribution at different depths and locations in patients. If patient dose protection is implemented, it is expected to shorten the test time to less than 5 minutes, which will bring significant efficiency to patient dose assurance. And quality improvement. This result shows that the detector is suitable for hospital daily proton beam warranty and patient dose warranty.
When the new proton therapy in Linkou Chang Gung Hospital uses a pen tip beam scan, the beam varies greatly in the lateral space. The detector can distinguish the slight difference in beam shape when the proton beam is incident at different positions. For more accurate calibration or to increase the beam dose rate, the detector can detect a rate of change of up to 10,000 times. At that time, the position and shape of the beam can be clearly recorded. According to this feature, the proton therapy device can be used in the future. The direction of quality assurance.
|
參考文獻 |
[1] Baskar, R., et al. "Cancer and radiation therapy: current advances and future directions." International journal of medical sciences 9.3 (2012): 193.
[2] Levin, W. P., et al. "Proton beam therapy." British journal of Cancer 93.8 (2005): 849.
[3] J.W. Boag, Ionization chambers, in: F.H. Attix, W.C. Roesch, E. Tochilin (Eds.) Radiation Dosimetry Vol. 2, Chap. 9, New York: Academic; 1966,1–72
[4] J.R. Greening Saturation characteristics of parallel-plate ionization chambers, Physics in Medicine & Biology, 9 (1964) 143 – 155.
[5] Lorentini, S., and M. Schwarz. "61. Dose-area-product determination with large planes-parallel ionization chambers: New approach to reference dosimetry in active scanning protontherapy." Physica Medica 56 (2018): 101.
[6] MarkFilipak, CC BY-SA 3.0, https://commons.wikimedia.org/w/index.php?curid=27983203
[7] Kohno, R., et al. "Development of continuous line scanning system prototype for proton beam therapy." International Journal of Particle Therapy 3.4 (2017): 429-438.
[8] ionization chamber. (n.d.) The Columbia Electronic Encyclopedia®. (2013). Retrieved March 17 2019 from https://encyclopedia2.thefreedictionary.com/ionization+chamber
[9] FlashForward Consortium , Varian Medical Systems, https://www.varian.com/oncology/solutions/proton-therapy/flashforward-consortium
[10] Flash Irradiation Delivered in a Clinical Treatment Room, IBA, https://iba-worldwide.com/content/pt/proton-flash-irradiation-delivered-clinical-treatment-room
[11] Lin, Y.C., et al. "Monte Carlo simulations for angular and spatial distributions in therapeutic-energy proton beams." Radiation Physics and Chemistry 140 (2017): 217-224.
[12] Hartmann, Bernadette, et al. "Investigations of a flat-panel detector for quality assurance measurements in ion beam therapy." Physics in Medicine & Biology 57.1 (2011): 51.
[13] Arjomandy, Bijan, et al. "Use of a two‐dimensional ionization chamber array for proton therapy beam quality assurance." Medical physics 35.9 (2008): 3889-3894.
[14] ICRU78, PRESCRIBING, RECORDING, AND REPORTING Proton Therapy, 2007.
[15] Hara, Yousuke, et al. "A Patient-Specific QA Procedure for Moving Target Irradiation in Scanned Ion Therapy." (2016): MOPB015.
[16] Zhu, X., et al. "Towards effective and efficient patient-specific quality assurance for spot scanning proton therapy." Cancers 7.2 (2015): 631-647.
[17] Schwaab, J., et al. "Experimental characterization of lateral profiles of scanned proton and carbon ion pencil beams for improved beam models in ion therapy treatment planning." Physics in Medicine & Biology 56.24 (2011): 7813.
[18] Actis, O., et al. "A comprehensive and efficient daily quality assurance for PBS proton therapy." Physics in Medicine & Biology 62.5 (2017): 1661.
[19] ICRU 50, PRESCRIBING, RECORDING, AND REPORTING Proton Therapy, 1993.
[20] ICRU 62, PRESCRIBING, RECORDING, AND REPORTING Proton Therapy, 1999.
[21] Vatnitskiy, S., P. Andreo, and D. T. L. Jones. "Recent advances in dosimetry in reference conditions for proton and light-ion beams." Proc. Int. Symp. on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (Vienna, 9–12 Nov. 2010). 2011.
[22] Low, Daniel A., and James F. Dempsey. "Evaluation of the gamma dose distribution comparison method." Medical physics 30.9 (2003): 2455-2464.
[23] Andreo, Pedro, et al. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. Vol. 398. IAEA technical report series, 2000.
[24] S.C. Lillicrap, et al. Code of practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service, Physics in Medicine & Biology, 35 (1990) 1355 – 1361.
[25] M. Liszka, et al. Ion recombination and polarity correction factors for a plane–parallel ionization chamber in a proton scanning beam, Medical Physics, 45 (2018) 391 – 401.
[26] J.B. Christensen, H. Tölli, N. Bassler, A general algorithm for calculation of recombination losses in ionization chambers exposed to ion beams. Medical physics, 43.10 (2016) 5484-5492.
[27] Lin, Yi-Chun, et al. "Monte Carlo simulations for angular and spatial distributions in therapeutic-energy proton beams." Radiation Physics and Chemistry 140 (2017): 217-224.
[28] R. Tansho, T. Furukawa, Y. Hara, et al. Experimental verification of gain drop due to general ion recombination for a carbon‐ ion pencil beam Experimental verification of gain drop due to general ion recombination for a carbon‐ion pencil beam, Medical Physics, 43 (2016) 635 – 642.
[29] Protection, Radiological. "ICRP publication 103." Ann ICRP 37.2.4 (2007): 2
[30] L. Vazquez-Quino, et al. "Patient specific pre-treatment QA verification using an EPID approach." Technology in cancer research & treatment 13.1 (2014): 1-10.
[31] 許琬婷,「鼻咽癌多種治療技術之放射治療計畫評量」,中臺科技大學,碩士論文,民國98年。
[32] 蔡綉吟,「質子束在水中橫向寬度及深度劑量曲線的量測與模擬」,中央大學,碩士論文,民國102年。
[33] Lovina Wijayanti,「Proton Energy Estimation by Least Square Method」,中央大學,碩士論文,民國106年。
[34] 彭瀚生,「多重線絲漂移室之初始性能」,中央大學,碩士論文,民國106年。
|