中大學術數位典藏-NCU Institutional Repository-提供博碩士論文、考古題、期刊論文、研究計畫等下載:Item 987654321/100868
English  |  正體中文  |  简体中文  |  全文笔数/总笔数 : 94274/94274 (100%)
造访人次 : 82851175      在线人数 : 1379
RC Version 7.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
搜寻范围 查询小技巧:
  • 您可在西文检索词汇前后加上"双引号",以获取较精准的检索结果
  • 若欲以作者姓名搜寻,建议至进阶搜寻限定作者字段,可获得较完整数据
  • 进阶搜寻
    NCU Institutional Repository > 理學院 > 化學學系 > 期刊論文 >  Item 987654321/100868


    jsp.display-item.identifier=請使用永久網址來引用或連結此文件: https://ir.lib.ncu.edu.tw/handle/987654321/100868


    题名: Determination of Dechlorane compounds in aqueous samples using ultrasound-assisted dispersive liquid-liquid microextraction and gas chromatography-electron-capture negative ion-mass spectrometry
    作者: 丁望賢;Hsieh, Huai-Kuang;Chen, Chien-Liang;Ding, Wang-Hsien
    贡献者: 理學院化學學系
    关键词: acetonitrile;carbon tetrachloride;centrifugation;chemical species;dispersants;flame retardants;glass;liquid-phase microextraction;sodium chloride;spectroscopy;standard deviation;ultrasonic treatment
    日期: 2013-12-02
    上传时间: 2026-04-21 14:16:41 (UTC+8)
    出版者: Royal Society of Chemistry
    摘要: 摘要: A simple and solvent-minimized procedure for the rapid determination of chlorinated flame retardants, Dechlorane compounds: syn- and anti-Dechlorane Plus (syn- and anti-DP), Dechlorane (Dec) 602, Dec 603 and Dec 604, in aqueous samples is described. The method involves the use of ultrasound-assisted dispersive liquid-liquid microextraction (UA-DLLME) prior to their determination by gas chromatography-electron-capture negative ion-mass spectrometry (GC-ECNI-MS) operating in the selected-ion-monitoring (SIM) mode. The parameters affecting the extraction efficiency of analytes from water samples were systematically investigated. The best extraction conditions involved the rapid injection of a mixture of 1.0 mL of acetonitrile (as a dispersant) and 60 mu L of carbon tetrachloride (as an extractant) into 10 mL of water sample containing 0.5 g of sodium chloride in a conical bottom glass tube. After ultrasonication for 1.0 min and centrifugation at 5000 rpm (5 min), the volume of the sedimented phase was adjusted to 20 mu L and the analytes were then determined by GC-ECNI-MS analysis. The limits of quantitation (LOQs) ranged from 0.08 to 0.3 ng L super(-1). The precision for these analytes, as indicated by relative standard deviations (RSDs), was less than 9% for both intra- and inter-day analyses. Accuracy, expressed as the mean extraction recovery, was between 75 and 92%. A preliminary result revealed that trace levels of Dechlorane compounds were detected in various water samples with total concentrations ranging from 0.6 to 0.9 ng L super(-1).
    出版日期: 2013-01-01
    出處: Analytical methods, 2013-01, Vol.5 (24), p.7001-7007
    資源來源: Royal Society of Chemistry
    識別號: ISSN: 1759-9660
    識別號: ISSN: 1759-9679
    識別號: EISSN: 1759-9679
    識別號: DOI: 10.1039/c3ay41505h
    显示于类别:[化學學系] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    index.html0KbHTML21检视/开启


    在NCUIR中所有的数据项都受到原著作权保护.

    社群 sharing

    ::: Copyright National Central University. | 國立中央大學圖書館版權所有 | 收藏本站 | 設為首頁 | 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::
    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 隱私權政策聲明