    题名: A noncontact detection technique for interfacial bone defects and osseointegration assessment surrounding dental implants
    作者: 潘敏俊;Zhuang, H.B.;Pan, M.Ch;Chen, J.Z.;Wu, J.W.;Chen, C.S.
    贡献者: 工學院機械工程學系
    关键词: Bone defect;Dental implant;Noncontact measurement;Osseointegration;Resonance frequency analysis
    日期: 2014-01-01
    上传时间: 2026-04-23 14:39:13 (UTC+8)
    出版者: Elsevier;Elsevier Ltd
    摘要: 摘要: •A noncontact detection technique developed to evaluate bone–implant interface imperfections.•In vitro (artificial bone defect models) and in vivo (animal models) trials were performed.•A prototyping design of an integrated excitation-and-detection transducer is proposed.•The examining technique and designated device are promising to assess dental implant stability. Dental implantation is popular treatment for edentulous and partially dentate patients. Effectively assessing interfacial bone defects and osseointegration requires considerable effort for successful implantation. This study presents the development of a noncontact detection technique that uses a loudspeaker and a noncontact cylindrical capacitive-type displacement sensor to measure and evaluate the severity and location of imperfections surrounding the bone–implant interface through resonance frequency (RF) analysis. To justify the effectiveness, both in vitro and in vivo experiments were performed. The acoustic sweep signal of a loudspeaker was used to excite models, and the vibration response signal from a displacement sensor was detected and transferred to the spectrum. The first RF values of the spectra on each model were subsequently identified and used to assess the interfacial situation. The measuring results show that the RF differences can be identified and applied to evaluate bone defects, as well as osseointegration. The proposed technique is a promising approach to aid dentists in assessing dental implant stability after surgery.
    出版者: Elsevier Ltd
    出版日期: 2014-09-01
    出處: Measurement : journal of the International Measurement Confederation, 2014-09, Vol.55, p.335-342
    版權: 2014 Elsevier Ltd
    識別號: ISSN: 0263-2241
    識別號: EISSN: 1873-412X
    識別號: DOI: 10.1016/j.measurement.2014.05.039