摘要(英) |
Dental implant after surgical implantation,Osteointegration occurs through the proliferation of bone cells on the surface of the implant, The interface tissue surrounding the implant provides stability, so the key to success after surgery is the degree of osseointegration, which is the stability of the implant. In recent years, this implantation method has been applied to limb amputations. After the amputation implant implantation operation, there are three stages of implant stability detection, which are the initial postoperative stability, the evaluation of the implant stability during rehabilitation, and the regular postoperative implant stability tracking.
The experimental planning is based on the osteoblast proliferation on the surface of the implant during the osseointegration to form the interface tissue to provide conditions for increasing the stability of the implant. This study plans to use epoxy resins with different Young′s coefficients as the interface between the femur and the implant. By not adding epoxy, pure epoxy, and epoxy + 30% glass fiber, corresponding to the three stages of the implant stability test, the bone interface tissue is simulated due to the rigidity change. The resonance frequency trend of the model can be used to evaluate whether the implant is stable State, based on which the finite element model and parameter settings are established.Applying ANSYS® 15.0 finite element analysis to estimate the possible vibration modes and resonance
frequencies. The geometric model of the implant is established with reference to the actual samples and 2D drawings provided by The Osseointegration Group of Australia. SolidWorks® is used to draw 3D files. The surface of the implant was originally a three-dimensional structure. Through the osseointegration process, the bone tissue filled with the three-dimensional structure was firmly connected to the inner wall of the medullary cavity.
Therefore, the three-dimensional surface was simplified to a fixed condition of planar contact when the implant model was established. Reverse engineering laser scanner was used to import, the material properties of each component were established in order by reference, and the model was cut and calculated using free mesh. In this study, epoxy resin was used to simulate the osseointegration interface tissue between the femur and the implant. When the osseointegration process is good, the interface tissue regeneration and remodeling make the femur and the implant tightly joint, and there is no friction between the femur, the implant, and the interface tissue. The finite element model sets the contact conditions of the interface tissue, and combines the two contact surfaces of the femur, the interface tissue, and the implant ; Without adding any interface tissue and its model contact conditions, in order to simulate the initial stability after surgery, bone cells have not yet regenerated and remodeled, and the contact surface between the femur and the implant has sliding and friction behaviors. For the friction state, the friction coefficient is defined as 0.58; because the bottom end of NCUpeg is screwed with the implant, it is regarded as fixed, so the contact surface conditions are set to the combined state. For numerical analysis, refer to the experimental vise clamping method, set the boundary conditions of the structural system to the fixed end of the cantilever beam structure, set the three most prominent positions of the femoral bottom as the fixed support points of the overall structure, perform modal analysis and discuss the resonance set 1-20 sets of modal shape results for review, select the modal shapes that match the experimental device to measure the same X and Y directions, and read the representative resonance frequency threshold value as a record. At the beginning of this study, Osstell®-SmartPeg was used as the magnetic accessory at the top of the implant. However, only high-frequency resonance frequencies were obtained during the experimental
measurement. According to the CAE modal analysis results, the high-frequency resonance frequency was only SmartPeg′s local vibration at the top of the implant. There is no vibration to the sheep bones and implants, which can not reflect the osseointegration status of lower limb implant surgery. This study designed a NCUpeg for lower limb implant detection. The experimental measurement results can obtain the low frequency resonance frequency. The low-frequency resonance frequency of the modal analysis is the displacement and swing of the entire sheep bone and the implant. The resonance frequency value can represent the stability of the lower limb implant implantation. Therefore, NCUpeg was selected as the magnetic accessory for this study. Using 4.0 mm as the segmentation, the analysis results of tissue resonance frequency thresholds of different interfaces were developed to evaluate the tendency of implant stability.
After experimental tests and finite element modal analysis and comparison verification, an NCUpeg is designed for lower limb implant detection. The modal analysis results are used to improve the magnetic accessories necessary for this measurement and prove the feasibility of NCUpeg in the measurement device. In this study, the results of the modal analysis of the resonance frequency data are integrated. If the stability of the implant gradually increases, the natural frequency will gradually increase. Therefore, the handheld detector can measure the tendency of the increase of the stability of the implant by the resonance frequency. |
參考文獻 |
[1] 張志涵、吳柱龍,「 個人化膝下義肢承套之製前生物力學評估」,行政院國家科學委員會專題研究計畫 成果報告(編號:NSC 94-2213-E-006-011),民國 95 年。
[2] 林銘川、陳思遠和廖珮茹,「 因周邊血管病變導致下肢截肢患者的長期存活及其義肢使用狀況之調查研究」,行政院國家科學委員會補助研究計畫 成果報告(編號:NSC 90-2314-B-002-309),民國 90 年。
[3] M. Al Muderis, W. Lu, J. J. Li, and B. Bosley, “Bone Mineral Density changes after Osseointegration,” ASAMI 2016 Brisbane, 1-31 (2016).
[4] J. Hughes, and N. A. Jacobs, “Prosthetics and Orthotics International,” The Journal of the
International Society for Prosthetics and Orthotics, 16(2), 1-71 (1992).
[5] 吳柱龍、李偉強和林耕任,「 評估膝下義肢合適性之系統建立—應用於糖尿病併發之截肢患者」,行政院國家科學委員會專題研究計畫 成果報告(編號:NSC 96-2218-E-142-001),民國 97 年。
[6] 重新輔具:產品介紹。20190102,https://www.chongxin.com.tw/products_73_75.htm
[7] R. Brånemark, Ö. Berlin, K. Hagberg, P. Bergh, B. Gunterberg, and B. Rydevik, “A novel osseointegrated percutaneous prosthetic system for the treatment of patients with transfemoral amputation,” The Bone & Joint Journal , 96-B(1), 1–16 (2014).
[8] L. Sennerby, and N. Meredith, “Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications,” Periodontology 2000, 47, 51-66 (2008).
[9] A. Khemka, L. Frossard, S. J. Lord, B. Bosley, and M. Al Muderis, “Osseointegrated total knee replacement connected to a lower limb prosthesis: 4 cases,” Acta Orthopaedica, 86(6), 1-16 (2015).
[10] K. Hagberg, and R. Brånemark, “One hundred patients treated with osseointegrated transfemoral amputation prostheses—Rehabilitation perspective,” Journal of Rehabilitation Research & Development, 46(3), 331-344 (2009).
[11] A. Thesleff, R. Brånemark, B. Håkansson, and M. O. Catala, “Biomechanical Characterisation of Bone-anchored Implant Systems for Amputation Limb Prostheses: A Systematic Review,” Ann Biomed Eng, 46(3), 377-391 (2018).
[12] Osseointegration Group of Australia : Spongiosa metal 。 20190105, https://www.osseointernational.com/background-of-osseointegration/history-of-osseointegration
[13] Osseointegration Group of Australia : What is Osseointegration? 。 20190110, https://www.osseointegration.org/osseointegration/
[14] W. Xu, and K. Robinson, “X-Ray Image Review of the Bone Remodeling Around an Osseointegrated Trans-femoral Implant and a Finite Element Simulation Case Study,” Annals of Biomedical Engineering, 36(3), 435-443 (2008).
[15] F. Shao, W. Xu, A. Crocombe, and D. Ewins, “Natural Frequency Analysis of Osseointegration for Trans-femoral Implant,” Annals of Biomedical Engineering, 35, 817–824 (2007).
[16] N. Meredith, “Assessment of Implant Stability as a Prognostic Determinant,” The International journal of Prosthodontics, 11, 491-501 (1998).
[17] M. Sjostrom, S. Lundgren, H. Nilson, and L. Sennerby, “Monitoring of implant stability in grafted bone using resonance frequency analysis A clinical study from implant placement to 6 months of loading,” Oral Maxillofacial Surgery, 34, 45-51 (2005).
[18] N. J. Cairns , M. J. Pearcy, J. Smeathers, and C. J. Adam, “Ability of modal analysis to detect osseointegration of implants in transfemoral amputees: a physical model study,” Med Biol Eng Comput, 51, 39-47 (2013).
[19] H. M. Huang , S. Y. Lee, C. Y. Yeh, and C. T. Lin, “Resonance frequency assessment of dental implant stability with various bone qualities: a numerical approach,” Clinical Oral Implants Research, 13, 65-74 (2002).
[20] 莊瀚伯,「牙科植體術後骨缺型態之結構分析」,國立中央大學機械工程學系,碩士論文,民國 95 年。
[21] W. C. C. Lee, J. M. Doocey, R. Brånemark, C. J. Adam, J. H. Evans, M. J. Pearcy, and L. A. Frossard, “FE stress analysis of the interface between the bone and an osseointegrated implant for amputees – Implications to refine the rehabilitation program,” Clinical Biomechanics, 23, 1243-1250 (2008).
[22] L. Zheng, J. M. Luo, B. C. Yang, J. Y. Chen, and X. D. Zhang, “3D Finite Element Analysis of Bone Stress around Distally Osteointegrated Implant for Artificial Limb Attachment,” Key Engineering Materials, 288-289, 653-656 (2005).
[23] M. Zhang, X. Dong, and Y. Fan, “Stress Analysis of Osseointegrated Transfemoral Prosthesis:A Finite Element Model,” Pro cesloodrys of 27th Annual Conference on Engineering in Medicine and Biology, 4060-4063 (2005).
[24] L. Meirovitch, Fundamentals of Vibrations., McGraw-Hill, New York (2001).
[25] N. Meredith, D. Alleyne, and P. Cawley, “Quantitative Determination of the Stability of the Implant‐tissue Interface Using Resonance Frequency Analysis,” Clin Oral Implants Res, 7(3), 261-267 (1996).
[26] 劉晉奇、褚晴暉,有限元素分析與 ANSYS 的工程應用,滄海,臺中 (2006)。
[27] 康淵、陳信吉,ANSYS 入門,臺北,全華(2008)。
[28] 元 智 大 學 徐 業 良 教 授 : 有 限 元 素 分 析 之 一 般 程 序 。20190120,https://www.youtube.com/watch?v=XPNFQaQ00ug
[29] H.H.Lee, Finite element simulations with ANSYS Workbench 12., McGraw-Hill., Taipei (2010).
[30] 謝易珊,「結構振型探討牙科植體術後臨床骨缺損」,國立中央大學機械工程學系,碩士論文,民國 101 年。
[31] 王栢村,電腦輔助工程分析之實務與應用,全華,臺北(2005)。
[32] 王栢村,振動學,全華,臺北(2002)。
[33] G. O. Glória, M. C. A. Teles, A. C. C. Neves, C. M. F. Vieira, F. P. D. Lopes, M. A. Gomes, F. M. Margem, and S. N. Monteiro, “Bending test in epoxy composites reinforced with continuous and aligned PALF fibers,” Journal of Materials Research and Technology, 306, 1-6 (2017).
[34] N. O. R. Maciel, J. B. Ferreira, J. S. Vieira, C. G. D. Ribeiro, F. P. D. Lopes, F. M. Margem, S. N. Monteiro, C. M. F. Vieira, and L. C. Silva, “Comparative tensile strength analysis between epoxy composites reinforced with curaua fiber and glass fiber,” Journal of Materials Research and Technology, 461, 1-5 (2018).
[35] B. M. Willie, R. D. Bloebaum, W. R. Bireley, K. N. Bachus, and A. A. Hofmann, “Determining relevance of a weight-bearing ovine model for bone ingrowth assessment,” Journal of Biomedical Materials Research, 69A, 567-576 (2004).
[36] T. J. Shelton, J. P. Beck, R. D. Bloebaum, and K. N. Bachus, “Percutaneous osseointegrated prostheses for amputees: Limb compensation in a 12-month ovine model,” Journal of Biomechanics, 44, 2601-2606 (2011).
[37] V. Varghese, “Finite element based design of hip joint prosthesis,” Rourkela National Institute of Technology Department of Biotechnology & Medical Engineering, Master of Technology, 1-47 (2011).
[38] T. Suzuki, I. Kanno, J. J. Loverich, H. Kotera, and K. Wasa, “Characterization of Pb(Zr,Ti)O3 thin films deposited on stainless steel substrates by RF-magnetron sputtering for MEMS applications,” Sensors and Actuators, A 125, 382-386 (2006).
[39] S. Checa, P. J. Prendergast, and G. N. Duda, “Inter-species investigation of the mechano-regulation of bone healing: Comparison of secondary bone healing in sheep and rat,” Journal of Biomechanics, 44, 1237-1245 (2011).
[40] Y. M. Lam, O. M. Pearson, C. W. Marean, and X. Chen, “Bone density studies in zooarchaeology,” Journal of Archaeological Science, 30, 1701-1708 (2003).
[41] S. Li, Y. Su, X. Zhu, H. Jin, Q. Ouyang, and D. Zhang, “Enhanced mechanical behavior and fabrication of silicon carbide particles covered by in-situ carbon nanotubes reinforced 6061 aluminum matrix composites.” Materials & Design, 107, 130-138 (2016).
[42] A. Shirazi-Adl, M. Dammak, and G. Paiement, “Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants.” Journal of Biomedical Materials Research, 27, 167-175 (1993).
[43] 李偉豪,「 手持式電磁型態植體骨整合檢測裝置暨驗證」, 國立中央大學機械工程學系,碩士論文,民國 109 年。
|