題名:
Tubular waveguide evanescent field absorption biosensor based on particle plasmon resonance for multiplex label-free detection
作者:
黃貞翰
;
Lin, Hsing-Ying
;
Huang, Chen-Han
;
Chen, Shin-Huei
;
Liu, Yu-Chia
;
Chang, Wei-Zhe
;
Chau, Lai-Kwan
貢獻者:
生醫理工學院生醫科學與工程學系
關鍵詞:
Absorption
;
antibodies
;
Biological and medical sciences
;
Biological Assay - instrumentation
;
Biosensing Techniques - instrumentation
;
Biosensor
;
Biosensors
;
Biotechnology
;
coatings
;
detection limit
;
Equipment Design
;
Equipment Failure Analysis
;
Fundamental and applied biological sciences. Psychology
;
High-throughput biochemical detection
;
Immunoassay - instrumentation
;
Light
;
Localized surface plasmon resonance
;
Methods. Procedures. Technologies
;
monitoring
;
Nanoparticle
;
Optical waveguide
;
refractive index
;
Refractometry - instrumentation
;
Reproducibility of Results
;
screening
;
Sensitivity and Specificity
;
Staining and Labeling
;
streptavidin
;
Surface Plasmon Resonance - instrumentation
;
urine
;
Various methods and equipments
日期:
2013-03-15
上傳時間:
2026-04-23 11:23:12 (UTC+8)
出版者:
Elsevier Ltd.;Kidlington: Elsevier B.V
摘要:
摘要: A novel tubular waveguide particle plasmon resonance (TW-PPR) sensor is demonstrated for label-free biochemical detection. The sensor itself is a microchamber of a defined sample volume, a mechanical support for sensor coating, a waveguide to provide evanescent wave interrogation, and it can be easily extended to a multi-channel format. The sensor resolution is estimated to be 2.6×10−6RIU in measuring solutions of various refractive indices. The sensing system can perform multiple measurements simultaneously and its limit of detection for anti-DNP antibody and streptavidin is 1.2×10−10g/ml (0.55pM) and 2.3×10−10g/ml (3.5pM), respectively. Accurate determination of these analytes with known concentration spiked in artificial urine were examined and the bias is less than ±7%, supporting the utility of the device for analyte screening in more complex media. The TW-PPR sensor can be inexpensively fabricated and has a special niche for monitoring biomolecular interactions in real-time, hence it is ideally suitable for disposable uses, especially promising for convenient high-throughput biochemical sensing applications. ► A tubular waveguide particle plasmon resonance sensor is used for label-free biochemical sensing. ► It relies on an unsophisticated light intensity interrogation scheme. ► A multi-channel sensing system is developed for higher-throughput biochemical detection.
其他題名: Biosens Bioelectron
出版者: Kidlington: Elsevier B.V
出版日期: 2013-03-15
出處: Biosensors & bioelectronics, 2013-03, Vol.41, p.268-274
版權: 2012 Elsevier B.V.
版權: 2014 INIST-CNRS
版權: Copyright © 2012 Elsevier B.V. All rights reserved.
識別號: ISSN: 0956-5663
識別號: ISSN: 1873-4235
識別號: EISSN: 1873-4235
識別號: DOI: 10.1016/j.bios.2012.08.041
識別號: PMID: 22964381
顯示於類別:
[生醫科學與工程學系] 期刊論文
文件中的檔案:
檔案
描述
大小
格式
瀏覽次數
index.html
0Kb
HTML
19
檢視/開啟
在NCUIR中所有的資料項目都受到原著作權保護.
社群 sharing
::: Copyright National Central University. | 國立中央大學圖書館版權所有 |
收藏本站
|
設為首頁
| 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::
DSpace Software
Copyright © 2002-2004
MIT
&
Hewlett-Packard
/
Enhanced by
NTU Library IR team
Copyright ©
-
隱私權政策聲明